ORIGINAL ARTICLE
Year : 2019  |  Volume : 112  |  Issue : 4  |  Page : 122-129

Ganglion cell analysis versus retinal nerve fiber layer thickness in glaucoma diagnosis


1 Department of Ophalmology, Banha University, Banha, Egypt
2 Department of Ophthalmology, Shibin Al Qanater Central Hospital, banha, Egypt

Date of Submission12-Sep-2019
Date of Acceptance24-Oct-2019
Date of Web Publication20-Feb-2020

Correspondence Address:
Sarah Mohamed Selim
MSc, Department of Ophthalmology, Faculty of Medicine, Banha University, Banha City, Qalubiah, 13511
Egypt
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ejos.ejos_45_19

Rights and Permissions
  Abstract 


Background Glaucoma is a progressive optic neuropathy, which can result in irreversible blindness. The main pathologic change in glaucoma is retinal ganglion cell loss. The macula has the greatest density of retinal ganglion cells (∼50%). The advent of spectral domain-optical coherence tomography (OCT) technology has allowed advanced macular imaging protocols to play an important role in the diagnosis and monitoring of glaucoma. The ganglion cell analysis is obtained by the Cirrus HD-OCT system segments and measures the thickness of the ganglion cell‑inner plexiform layer, thereby potentially increasing its diagnostic accuracy compared with conventional peripapillary retinal nerve fiber layer (RNFL) thickness measurement.
Patients and methods The study was done on 40 eyes of 40 patients, and patients were divided into two groups: Group 1 included 20 adult patients more than 18 years who were diagnosed with primary open-angle glaucoma. Group 2 included 20 normal controls. One eye of each individual was prospectively enrolled. The OCT was done using 3D-OCT 2000 (Topcon) to evaluate the following: RNFL parameters (average total thickness, superior average thickness, and inferior average thickness) and ganglion cell complex (GCC) parameters (total average thickness, superior average thickness, and inferior average thickness).
Result There was a positive statistically significant correlation of high probability between the two groups regarding RNFL thickness (P<0.001). There was a positive statistically significant correlation of high probability between the two groups regarding GCC thickness (P<0.001). There was a positive statistical significant correlation of high probability between the groups regarding intraocular pressure (P<0.001). There was a positive statistical significant correlation of high probability between the groups regarding cup-to-disc ratio (P<0.001).
Conclusion Early diagnosis of glaucoma and early initiation of treatment is extremely important, as further vision loss can be stopped or slowed down. RNFL and GCC measurement with spectral domain-OCT could provide important information for detection and evaluation of glaucoma. There is strong positive correlation between the RNFL thickness and the GCC thickness in the glaucomatous patients.

Keywords: ganglion cell, glaucoma, optical coherence tomography, retinal nerve fiber


How to cite this article:
Soliman TT, Gad EA, Selim SM. Ganglion cell analysis versus retinal nerve fiber layer thickness in glaucoma diagnosis. J Egypt Ophthalmol Soc 2019;112:122-9

How to cite this URL:
Soliman TT, Gad EA, Selim SM. Ganglion cell analysis versus retinal nerve fiber layer thickness in glaucoma diagnosis. J Egypt Ophthalmol Soc [serial online] 2019 [cited 2020 Apr 7];112:122-9. Available from: http://www.jeos.eg.net/text.asp?2019/112/4/122/278810




  Introduction Top


Glaucoma is defined as a progressive optic neuropathy, which can result in irreversible blindness. The big challenge facing ophthalmologists is the early diagnosis of the disease and how to prevent its progression [1].

The main pathologic change in glaucoma is retinal ganglion cell (RGC) loss, which results in atrophy of all related inner retinal layers, namely, retinal nerve fiber layer (RNFL), which are axons of ganglion cells (GC); ganglion cell layer (GCL), which has the body of GCs; and inner plexiform layer (IPL), having the dendrites of GCs [1].

There is a good correlation found between RNFL thickness attenuation and visual field loss [2].

The macula was found to have the greatest number of RGCs (∼50%) [3].

Objective quantitative analysis of RNFL thickness has become the standard method for the detection of glaucoma since the introduction of optical coherence tomography (OCT) [3].

Macular thickness measurements with the Stratus OCT was not found to be superior to RNFL measurements [2].

The use of spectral domain-optical coherence tomography (SD-OCT) technology has allowed macular imaging to play an important role in the diagnosis and monitoring of glaucoma [4]. The ganglion cell analysis (GCA) is obtained by the Cirrus HD-OCT system (Topcon) segments and measures the thickness of the GC‑IPL, so increasing its diagnostic accuracy compared to conventional RNFL thickness measurement [5].

Published studies have reported comparing values of GCA and RNFL thickness for diagnosing glaucoma [5].

If there is a visual field defect, a differentiation between glaucomatous patients and normal is not difficult. The challenge comes in how to differentiate a glaucoma suspect from normal when visual field is normal [6].

If there is a significantly lesser number of RGCs in this group of patients, it may result in earlier intervention in the disease before functional loss happens [7].

Although RNFL thickness measurement has proven to be useful in the diagnosis of glaucoma, as RNFL loss is preceded by RGC death, so measurement of this layer (RGC) helps to detect the disease early [8].


  Patients and methods Top


An high selected cross sectional study was granted the prerequisite acceptance from the Medical Bioethics Commity of Faculty of Medicine Benha University from July 2018 to December 2018.

Patients were divided into two groups:
  • Group 1 included 20 adult patients (>18 years) who are diagnosed with primary open-angle glaucoma (POAG).
  • Group 2 included 20 normal controls.


One eye of each individual was prospectively enrolled. Informed consent was taken from all participants.

Inclusion criteria

Adult patients, above 18 years, already diagnosed with POAG, good visual acuity, with clear media were included.

Control group included adult individuals above 18 years, with normal intraocular pressure (IOP) and normal visual field, having no history of any eye disease.

Exclusion criteria

Patients with media opacity interfering with good-quality OCT scans such as corneal opacity, cataracts, etc.; history of intraocular diseases; complicated intraocular surgery; nonglaucomatous secondary causes of elevated IOP (e.g. iridocyclitis and trauma); coexisting retinal disease (e.g. diabetic retinopathy); other diseases affecting visual field (e.g. pituitary lesions); taking medications known to affect visual field sensitivity; or with problems other than glaucoma affecting color vision were excluded from this study.

Tests

Each patient will have a comprehensive ophthalmic examination, including best-corrected visual acuity, IOP) measured by Goldmann applanation tonometry, slit-lamp biomicroscopy, and stereoscopic fundus evaluation on the slit lamp using a 90.0 D lens optic discs.

All included individuals were scanned with OCT.

Optical coherence tomography scanning

For RNFL thickness measurements, the optic disc 200×200 scan will be used to acquire a cube of side 6 mm, while the patient is fixated so that the optic disc will be near the center of the scan.

For macular GCA, the macular cube 200×200 protocol will be used for GCA.

Optical coherence tomography macular scan and optic nerve head (ONH) scan were performed using the Topcon SD-OCT model 2000 version.

All patients signed an informed consent under the tenets of the Helsinki guidelines.

Statistical analysis

The results were analyzed using the SPSS for Windows software, version 10.0, (SPSS Inc., Chicago, Illinois, USA). Descriptive statistics were computed for all the variables measured.


  Results Top


Sex

Our study included 40 eyes of 40 patients divided into two groups:
  1. Group 1 included 20 patients with POAG, comprising 11 (55%) males and nine (45%) females.
  2. Group 2 included 20 normal controls, comprising 11 (55%) males and nine (45%) females, as shown in [Table 1] and [Figure 1].
    Table 1 Comparison between the two studied groups according to sex

    Click here to view
    Figure 1 Sex of two studied groups.

    Click here to view


There was an insignificant correlation between the two groups according to sex, as shown in [Figure 1].

Group 1: patients with primary open-angle glaucoma

Of 20 patients with POAG, 11 were on one line of treatment of antiglaucoma drops and nine were on two line of treatments, as shown in [Table 2] and [Table 3] and [Figure 2] and [Figure 3].
Table 2 Distribution of the studied cases according to number of line of treatment

Click here to view
Table 3 Distribution of the studied cases according to duration of the disease

Click here to view
Figure 2 Distribution of the studied cases according to treatment.

Click here to view
Figure 3 Distribution of the studied cases according to duration of disease.

Click here to view


Age

The mean age of the patients in group 1 was 54.60±10.67 years, ranging from 37 to 71 years. Group 2 had mean age of 40.75±13.55 years, ranging from 20 to 67 years ([Table 4] and [Figure 4]). There was a positive statistically significant correlation of high probability between the group 1 and group 2 regarding age (P=0.001).
Table 4 Age of the studied groups

Click here to view
Figure 4 Age of studied groups.

Click here to view


Visual acuity

As shown in [Table 5] and [Figure 5], group 1 had mean V/A of 0.52±0.26 logMar, ranging from 0.2 to 1, and group 2 had mean V/A of 0.39±0.20 logMar, ranging from 0.0 to 0.8.
Table 5 Comparison between the two studied groups according to best-corrected visual acuity

Click here to view
Figure 5 Comparison between the two studied groups according to BCVA. BCVA, best-corrected visual acuity.

Click here to view


There was an insignificant correlation of low probability between the both groups regarding best-corrected visual acuity (P=0.127).

Intraocular pressure

Group 1 had mean IOP of 18.40±4.08, ranging from 12 to 28 mmHg and group 2 had mean IOP of 13.20±1.88, ranging from 10 to16 mmHg, as seen in [Table 6] and [Figure 6]. There was a positive statistically significant correlation of high probability between both groups regarding IOP (P<0.001).
Table 6 Comparison between the two studied groups according to intraocular pressure

Click here to view
Figure 6 Comparison between the two studied groups according to IOP. IOP, intraocular pressure.

Click here to view


Cup-to-disc ratio

[Table 7] and [Figure 7] show that in group 1 had an average cup-to-disc (C/D) area ratio of 0.63±0.10, ranging from 0.5 to 0.8 and group 2 had an average C/D area ratio of 0.46±0.06, ranging from 0.4 to 0.6.
Table 7 Comparison between the two studied groups according to cup-to-disc ratio

Click here to view
Figure 7 Comparison between the two studied groups according to C/D ratio. C/D, cup-to-disc.

Click here to view


There was a positive statistically significant correlation of high probability between both groups regarding C/D ratio (P<0.001).

Optical coherence tomography findings

Retinal nerve fiber layer parameters

Group 1 had an average RNFL thickness of 25.55±3.97 μm, ranging from 19.0 to 32.0 μm. The superior average thickness was 25.35±6.31 μm, ranging from 18.0 to 38.0 μm, and the inferior average thickness was 26.35±4.58 microns, ranging from 20.0 to 33.0.

Group 2 had an average RNFL thickness of 56.95±9.68 μm, ranging from 44.0 to 74.00 μm. The superior average thickness was 55.95±9.43 μm, ranging from 44.0 to 71.00 μm, and the inferior average thickness was 58.30±10.15 μm, ranging from 44.0 to 77.0 μm, as shown in [Table 8] and [Figure 8].
Table 8 Comparison between the two studied groups according to retinal nerve fiber layer

Click here to view
Figure 8 Comparison between the two studied groups according to RNFL. RNFL, retinal nerve fiber layer.

Click here to view


There was a positive statistically significant correlation of high probability between the two groups regarding RNFL thickness (P<0.001).

The ganglion cell complex parameters

  1. Group 1 had an average ganglion cell complex (GCC) thickness of 69.35±12.78 μm, ranging from 52.0 to 93.0 μm. The superior average thickness of the GCC was 69.70±16.48 μm, ranging from 48.0 to 107.0 μm, and the inferior average thickness was 69.45±10.75 μm, ranging from 57.0 to 95.0 μm.
  2. Group 2 had an average GCC thickness of 111.6±9.37 μm, ranging from 99.0 to 135.0 μm. The superior average thickness of the GCC was 112.8±10.30 μm, ranging from 100.0 to 148.0 μm, and the inferior average thickness was 110.6±8.83 μm, ranging from 99.0 to 132.0 μm, as seen in [Table 9] and [Figure 9].
    Table 9 Comparison between the two studied groups according to ganglion cell complex thickness

    Click here to view
    Figure 9 Comparison between the two studied groups according to GC. GC, ganglion cells.

    Click here to view


There was a positive statistically significant correlation of high probability between the two groups regarding GCC thickness (P<0.001).


  Discussion Top


Glaucoma occurs owing to death of RGCs, and RGCs being concentrated at the macula lead to the use of macular thickness as a useful discriminator for glaucoma. Although the GCA and RNFL are measured at different locations of the retina, as the macula contains 50% of RGCs in the retina, measuring macular GCL appears to be a good approach to assess GC death. The RNFL measured around the disc are axons of GCs which converge on the optic disc before leaving the eye [1].

A study by Inuzuka et al. [9] observed a significant correlation between the macular GCC of the inner or outer sector of the parafovea and in the change of the visual field in each hemifield defect or the apparently normal hemifield. The decrease of the GCC corresponding to the apparently normal hemifield correlated with the progression of the severity of the glaucomatous defects.

In a study by Takagi et al. [10], it was found that the thickness of the macular GCC in the normal hemifield of the glaucomatous eyes was significantly less than in normal eyes. In contrast, the total thickness of the macular retinas between the glaucomatous and normal eyes showed no significant difference. In addition, the thicknesses of the macular GCC and RNFL in the normal hemisphere of the glaucomatous eyes significantly correlated with the total deviation in the visual field parameters of the corresponding area.

In a study by Na et al. [11], it was found that perimetrically normal hemifields of glaucomatous eyes had significantly lower macular GCC and pRNFL thicknesses than did the corresponding retinal regions of healthy eyes. SD-OCT may be a diagnostic tool for evaluation of early macular and circumpapillary structural changes in glaucomatous eyes with localized VF defects.

A study by Moreno et al. [12] found that the ability of macular GC–IPL parameters to discriminate normal eyes and eyes with early glaucoma is high and comparable to that of the best pRNFL and ONH parameters.

Sevim et al. [13] in their study found that GCC and RNFL thickness measurements performed by FD-OCT showed high diagnostic ability in detecting glaucoma.

In the study by Oddone et al. [14], with the use of OCT, RNFL parameters are still preferable to macular parameters for diagnosing manifest glaucoma, but the differences are small.

A study of Schulze et al. [15] found that imaging of the GCC using FD-OCT (RTVue-100) has a comparable diagnostic ability to RNFL and ONH measurements in distinguishing between patients with glaucoma and healthy participants. No differences were found between patients with OHT and normal participants regarding ONH, RNFL, and GCC parameters.

Moreover, Medeiros et al. [16] in their study found that glaucomatous eyes with the earliest detectable visual field loss on automated perimetry may already show substantial loss of RGCs. Empirical estimates of RGC counts combining structural and functional tests agreed closely with previous histologic reports on the number of RGCs associated with early visual fields defects.

Our study found that there was a positive statistically significant correlation of high probability between the two groups regarding RNFL thickness (P<0.001)

Group 1 had an average RNFL thickness of 25.55±3.97 μm, ranging from 19.0 to 32.0 μm. The superior average thickness was 25.35±6.31 μm, ranging from 18.0 to 38.0 μm. The inferior average thickness was 26.35±4.58 μm, ranging from 20.0 to 33.0.

Group 2 had an average RNFL thickness of 56.95±9.68 μm, ranging from 44.0 to 74.00 μm. The superior average thickness was 55.95±9.43 μm, ranging from 44.0 to 71.00 μm. The inferior average thickness was 58.30±10.15 μm, ranging from 44.0 to 77.0 μm ([Table 8]).

There was a highly statistically significant correlation between the two groups in GCC thickness (P<0.001).
  1. Group 1 had an average GCC thickness of 69.35±12.78 μm, ranging from 52.0 to 93.0 μm. The superior average thickness of the GCC was 69.70±16.48 μm, ranging from 48.0 to 107.0 μm. The inferior average thickness was 69.45±10.75 μm, ranging from 57.0 to 95.0 μm.
  2. Group 2 had an average GCC thickness of 111.6±9.37 μm, ranging from 99.0 to 135.0 μm. The superior average thickness of the GCC was 112.8±10.30 μm, ranging from 100.0 to 148.0 μm. The inferior average thickness was110.6±8.83 μm, ranging from 99.0 to 132.0 μm ([Table 9]).


Single baseline measurements of GCC layer in glaucomatous and nonglaucomatous eyes also showed significant difference (P<0.001 for superior, inferior, and total GCC values in normal eyes compared with glaucomatous eyes). The total GCC values in glaucomatous eyes compared with normal eyes showed very significant statistical difference.


  Conclusion Top


Diagnosis of glaucoma and early initiation of treatment is extremely important, as further vision loss can be stopped or slowed down. RNFL and GCC measurement with SD-OCT could provide important information for detection and evaluation of glaucoma. There is a strong positive correlation between the RNFL thickness and the GCC thickness in the glaucomatous patients. Imaging of the GCC has a comparable diagnostic ability to RNFL and ONH measurements in distinguishing between patients with glaucoma and healthy participants.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.



 
  References Top

1.
Weinreb RN, Aung T, Medeiros FA. The pathophysiology and treatment of glaucoma: a review. JAMA 2014; 311:1901–1911.  Back to cited text no. 1
    
2.
Ajtony C, Balla Z, Somoskeoy S, Kovacs B. Relationship between visual field sensitivity and retinal nerve fiber layer thickness as measured by optical coherence tomography. Invest Ophthalmol Vis Sci 2007; 48:258–263.  Back to cited text no. 2
    
3.
Greenfield DS, Bagga H, Knighton RW. Macular thickness changes in glaucomatous optic neuropathy detected using optical coherence tomography. Arch Ophthalmol 2003; 121:41–46.  Back to cited text no. 3
    
4.
Sung KR, Kim JS, Wollstein G, Folio L, Kook MS, Schuman JS. Imaging of the retinal nerve fibre layer with spectral domain optical coherence tomography for glaucoma diagnosis. Br J Ophthalmol 2011; 95:909–914.  Back to cited text no. 4
    
5.
Mwanza JC, Oakley JD, Budenz DL, Chang RT, O’Rese JK, Feuer WJ. Macular ganglion cell-inner plexiform layer: automated detection and thickness reproducibility with spectral domain-optical coherence tomography in glaucoma. Invest Ophthalmol Vis Sci 2011; 52:8323–8329.  Back to cited text no. 5
    
6.
Tan O, Li G, Lu AT, Varma R, Huang D, Advanced Imaging for Glaucoma Study Group. Mapping of macular substructures with optical coherence tomography for glaucoma diagnosis. Ophthalmology 2008; 115:949–956.  Back to cited text no. 6
    
7.
Weinreb RN, Khaw PT. Primary open-angle glaucoma. Lancet 2004; 363:1711–1720.  Back to cited text no. 7
    
8.
8.Morooka S, Hangai M, Nukada M, Nakano N, Takayama K, Kimura Y et al. Wide 3‑dimensional macular ganglion cell complex imaging with spectral‑domain optical coherence tomography in glaucoma. Invest Ophthalmol Vis Sci 2012; 53:4805–4812.  Back to cited text no. 8
    
9.
Inuzuka H, Kawase K, Yamada H, Oie S, Kokuzawa S, Yamamoto T. Macular ganglion cell complex thickness in glaucoma with superior or inferior visual hemifield defects. J Glaucoma 2014; 23:145–149.  Back to cited text no. 9
    
10.
Takagi ST, Kita Y, Yagi F, Tomita G. Macular retinal ganglion cell complex damage in the apparently normal visual field of glaucomatous eyes with hemifield defects. J Glaucoma 2012; 21:318–325.  Back to cited text no. 10
    
11.
Na JH, Kook MS, Lee Y, Yu SJ, Choi J. Detection of macular and circumpapillary structural loss in normal hemifield areas of glaucomatous eyes with localized visual field defects using spectral-domain optical coherence tomography. Graefes Arch Clin Exp Ophthalmol 2012; 250:595–602.  Back to cited text no. 11
    
12.
Moreno PA, Konno B, Lima VC, Castro DP, Castro LC, Leite MT et al. Spectral-domain optical coherence tomography for early glaucoma assess ment: analysis of macular ganglion cell complex versus peripapillary retinal nerve fiber layer. Can J Ophthalmol 2011; 46:543–547.  Back to cited text no. 12
    
13.
Sevim MS, Buttanri B, Acar BT, Kahya A, Vural ET, Acar S. Ability of Fourier-domain optical coherence tomography to detect retinal ganglion cell complex atrophy in glaucoma patients. J Glaucoma 2013; 22:542–549.  Back to cited text no. 13
    
14.
Oddone F, Lucenteforte E, Michelessi M, Rizzo S, Donati S, Parravano M, Virgili G. Macular versus retinal nerve fiber layer parameters for diagnosing manifest glaucoma: a systematic review of diagnostic accuracy studies. Ophthalmology 2016; 123:939–949.  Back to cited text no. 14
    
15.
Schulze A, Lamparter J, Pfeiffer N, Berisha F, Schmidtmann I, Hoffmann EM. Diagnostic ability of retinal ganglion cell complex, retinal nerve fiber layer, and optic nerve head measurements by Fourier-domain optical coherence tomography. Graefes Arch Clin Exp Ophthalmol 2011; 249:1039–1045.  Back to cited text no. 15
    
16.
Medeiros FA, Lisboa R, Weinreb RN, Liebmann JM, Girkin C, Zangwill LM. Retinal ganglion cell count estimates associated with early development of visual field defects in glaucoma. Ophthalmology 2013; 120:736–744.  Back to cited text no. 16
    


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7], [Figure 8], [Figure 9]
 
 
    Tables

  [Table 1], [Table 2], [Table 3], [Table 4], [Table 5], [Table 6], [Table 7], [Table 8], [Table 9]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Patients and methods
Results
Discussion
Conclusion
References
Article Figures
Article Tables

 Article Access Statistics
    Viewed172    
    Printed20    
    Emailed0    
    PDF Downloaded38    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]